The Stator Water-cooled Generator Insulation Characteristic Tester is specially used for measurement tests of water-cooled generators, and can also be used for insulation tests in laboratories or on-site. This instrument has the function of automatically compensating and eliminating the influence of polarization potential on measurement results, fundamentally overcoming the defects of other similar products.
The Stator Water-cooled Generator Insulation Characteristic Tester has strong load capacity, outputs rated negative high voltage of 2500V, short-circuit current about 40mA; when load is 70kΩ, voltage drop <10%. In addition to accurately measuring insulation performance, this instrument can also automatically measure temperature, humidity, and time; automatically calculate absorption ratio and polarization index; can store each measurement result; comes with a built-in printer; can communicate with a computer.
2. The operator of the instrument should have general knowledge of electrical equipment or instruments.
3. This instrument can be used indoors and outdoors, but should avoid places with rain, corrosive gases, excessive dust, high temperature, direct sunlight, etc.
4. The instrument should avoid severe vibration.
5. Maintenance, care, and adjustment of the instrument should be performed by professionals.
6. It is strictly forbidden to touch the test leads during testing!
7. After testing, it is best to short-circuit the test terminals for manual discharge!
8. Strictly comply with safety operating procedures.
9. Non-test personnel must stay away from the high-voltage test area, and the test area must be clearly marked with fences or ropes, warning signs, etc.